ACL ELITE Series
Coagulology,

Параметр Значение Ед. измерения
Основные параметры
Auto shutdown/Auto startup programmable not needed/not needed
Auto. detects adequate reagents for aspiration and analysis yes
Autocalib. or autocalib. alert/Multipoint calib. supported no/yes
Automatic rerun capability/Auto reflex testing capability yes/yes
Automatic transfer of QC results to LIS yes
Bidirectional interface capability yes (broadcast download and host query)
Clot detection as preanalytical variable in plasma sample no
Country where analyzer designed/Manufactured U.S./U.S.
Data-management capability yes
Dilution of patient samples onboard yes
Dimensions (H ? W ? D)/Weight/Instrument footprint 24 ? 37 ? 24 inches/139 lbs/6 square feet
Disposables used/Price of each cuvettes/varies
Distinguishing features (supplied by company) test menu featuring D-Dimer; barcoded reagent management; ACL family harmonization; HemosIL INR plasma sets for INR test system validation or calibration, or both; HemosIL liquid Anti-Xa with universal calibration curve for UFH and LMWH
Electronic interface available (or will be) to automated (or robotic) specimen handling system no
Factor assays require manual manipulation or dilutions no
FDA-cleared chromogenic tests Anti-Xa
FDA-cleared clotting-based tests PT, APTT, fibrinogen, TT, factors
FDA-cleared immunologic tests D-dimer, VWF (Act. and Ag.), free protein S, factor XIII Ag., homocysteine
First year sold 2006
Hemolysis/Turbidity detection-quantitation no/no
How labs get LOINC codes for reagent kits
Interface supplied by instrument vendor no
Interfaces in active user sites for: most major vendors
Lag time during which hypercoagulable sample not detected yes (PT and PTT: 3 seconds)
List price
LOINC codes transmitted with all results no
Maximum time same lot number of reagents can be used 18 months
Measures No. of tests remaining/Short sample detection yes/yes
Methodologies supported clot detection, LED optical (nephelometric), chromogenic, immunologic
Minimum sample volume aspirated precisely at one time 5 µL
Model type benchtop
Modem servicing no
Multiple reagent configurations supported yes
Number of different assays programmed and calib. at one time 300
Number of different measured assays onboard simultaneously 22
Number of reagent containers onboard at once/Tests per container/Reagents refrigerated onboard 22/varies/yes
Number of units installed in U.S./Outside U.S. 4,000+/8,000+ (all ACL models combined)
Number of user-definable (open) channels 100
Of those defined, number active simultaneously 20
Onboard maintenance records yes
Onboard test automatic inventory yes
Operates on whole blood or spun plasma spun plasma
Operational type modified random access
Operator must load sep. reagent pack per specimen/Test run no/no
Primary tube sampling supported/Pierces caps on primary tubes yes/no
Producer Instrumentation Laboratory
Read time extended for prolonged clotting times yes
Reagent barcode reading capability yes
Reagent type open reagent system
Reagents, consumables loaded without interrupting testing yes
Results transferred to LIS as soon as test time complete yes
Same capabilities when third-party reagent used yes
Sample barcode reading capability yes
Sample handling system tray, primary tubes
Standard specimen volume required to run PT or PTT/ Factor VIII activity 60 ?L/18 ?L
Stat time to complete all analytes/Throughput per hour for Factor VIII activity assay 8 minutes/125 specimens
Stat time to complete all analytes/Throughput per hour for Fibrinogen 4 minutes/175 specimens
Stat time to complete all analytes/Throughput per hour for PT alone 4 minutes/175 specimens
Stat time to complete all analytes/Throughput per hour for PT, PTT 8 minutes/270 specimens
Supports direct-from-track sampling no
Tests in development but not yet submitted
Time delay from ordering stat to aspiration of sample 15 seconds
Time required for maintenance by lab personnel daily: <5 minutes; weekly: 10 minutes; monthly: 5 minutes
User can set different-than-standard Incubation times/Reading times yes/yes
User can set different-than-standard No. and sources of reagent yes
User can set different-than-standard Reagent volumes/Sample volumes yes/yes
Walkaway capacity: Number of specimens/Number of tests 40/260

;