• News показать все
  • [30.09.2017] Клиническая значимость белка B7-H4 в качестве нового прогностического маркера рака молочной железы
  • [30.09.2017] Клинический термоанализатор позволяет проводить ряд лабораторных тестов
  • [30.09.2017] Масс-спектрометрическое обнаружение становится более масштабным

Sysmex CA-1500
Coagulology,

Параметр Значение Ед. измерения
Основные параметры
Auto shutdown/Auto startup programmable no/no
Auto. detects adequate reagents for aspiration and analysis yes
Autocalib. or autocalib. alert/Multipoint calib. supported no/yes
Automatic rerun capability/Auto reflex testing capability yes/yes
Automatic transfer of QC results to LIS yes
Bidirectional interface capability yes (host query)
Clot detection as preanalytical variable in plasma sample no
Country where analyzer designed/Manufactured Japan/Japan
Data-management capability onboard (includes QC: L-J plots and Westgard multirules)
Dilution of patient samples onboard yes
Dimensions (H ? W ? D)/Weight/Instrument footprint 20 ? 31.2 ? 31.2 inches/186 lbs/6.8 square feet
Disposables used/Price of each reaction tubes, sample plates, CA clean I and II, system buffer, halogen lamp, closed container sample replacement needles/varies with volume
Distinguishing features (supplied by company) simultaneous curve calibrating and patient testing; ability to load multiple bottles or multiple lots of reagent; user-definable, repeat, redilute, and reflex testing
Electronic interface available (or will be) to automated (or robotic) specimen handling system yes (Sysmex CST series)
Factor assays require manual manipulation or dilutions no
FDA-cleared chromogenic tests Innovance AT, Berichrom AT, plasminogen, factor VIII chromo, alpha-2 antiplasmin, protein C chromo, heparin
FDA-cleared clotting-based tests PT, APTT, fibrinogen, TT, reptilase time, factor assays, DRVVT (screen and confirm), factor V Leiden, protein C clot, protein S activity
FDA-cleared immunologic tests Innovance D-dimer
First year sold U.S.: 2000; worldwide: 1999
Hemolysis/Turbidity detection-quantitation no/yes
How labs get LOINC codes for reagent kits
Interface supplied by instrument vendor no
Interfaces in active user sites for: all major LIS vendors
Lag time during which hypercoagulable sample not detected yes (PT: 7 seconds, APTT: 15 seconds)
List price $97,529 standard model; $110,544 cap-piercing model
LOINC codes transmitted with all results no
Maximum time same lot number of reagents can be used 1 year
Measures No. of tests remaining/Short sample detection yes/yes
Methodologies supported clot detection, optical; turbidimetric, chromogenic, immunologic
Minimum sample volume aspirated precisely at one time 5 µL
Model type benchtop
Modem servicing no
Multiple reagent configurations supported yes
Number of different assays programmed and calib. at one time 25
Number of different measured assays onboard simultaneously 15
Number of reagent containers onboard at once/Tests per container/Reagents refrigerated onboard 39/up to 200/yes (15°C)
Number of units installed in U.S./Outside U.S.
Number of user-definable (open) channels 25
Of those defined, number active simultaneously 15
Onboard maintenance records no
Onboard test automatic inventory yes
Operates on whole blood or spun plasma spun plasma
Operational type continuous random access
Operator must load sep. reagent pack per specimen/Test run no/no
Primary tube sampling supported/Pierces caps on primary tubes yes (3–5 mL)/yes
Producer Siemens Healthcare Diagnostics
Read time extended for prolonged clotting times yes (selectable on menus)
Reagent barcode reading capability yes
Reagent type open reagent system (lyophilized, reconstituted manually), optimized for Siemens instruments
Reagents, consumables loaded without interrupting testing some consumables yes, reagents no
Results transferred to LIS as soon as test time complete yes
Same capabilities when third-party reagent used yes
Sample barcode reading capability yes
Sample handling system 10-tube position sample rack ? 5
Standard specimen volume required to run PT or PTT/ Factor VIII activity 50 µL/5 µL
Stat time to complete all analytes/Throughput per hour for Factor VIII activity assay 8 minutes/—
Stat time to complete all analytes/Throughput per hour for Fibrinogen 8 minutes/120 results
Stat time to complete all analytes/Throughput per hour for PT alone 7 minutes/120 results
Stat time to complete all analytes/Throughput per hour for PT, PTT 8 minutes/80 results
Supports direct-from-track sampling yes (Sysmex CST series)
Tests in development but not yet submitted
Time delay from ordering stat to aspiration of sample 2 minutes
Time required for maintenance by lab personnel daily: <5 minutes; weekly: 1 minute; quarterly: <5 minutes
User can set different-than-standard Incubation times/Reading times yes/yes
User can set different-than-standard No. and sources of reagent yes
User can set different-than-standard Reagent volumes/Sample volumes yes/yes
Walkaway capacity: Number of specimens/Number of tests 50/up to 1,000

;