Sysmex CA-600 systems
Coagulology,

Параметр Значение Ед. измерения
Основные параметры
Auto shutdown/Auto startup programmable no/no
Auto. detects adequate reagents for aspiration and analysis yes
Autocalib. or autocalib. alert/Multipoint calib. supported no/yes
Automatic rerun capability/Auto reflex testing capability no/no
Automatic transfer of QC results to LIS yes
Bidirectional interface capability yes (host query)
Clot detection as preanalytical variable in plasma sample no
Country where analyzer designed/Manufactured Japan/Japan
Data-management capability onboard (includes QC: L-J plots)
Dilution of patient samples onboard yes
Dimensions (H ? W ? D)/Weight/Instrument footprint 22.5 ? 19.5 ? 19.5 inches/~94.6 lbs/~3.05 square feet
Disposables used/Price of each reaction tubes, CA-clean I, CA-clean II, thermal paper/varies with volume
Distinguishing features (supplied by company) maximizes counter space w/compact footprint in low-volume labs; increases uptime/reduces service expenses; two models: CA-620 for routine clotting-based testing, CA-660 for clotting, chromogenic, immunologic testing needs
Electronic interface available (or will be) to automated (or robotic) specimen handling system no
Factor assays require manual manipulation or dilutions no
FDA-cleared chromogenic tests Innovance AT, Berichrom AT, protein C chromo, heparin
FDA-cleared clotting-based tests PT, APTT, fibrinogen, TT, reptilase time, protein C clot, factor assays
FDA-cleared immunologic tests Innovance D-dimer
First year sold 2012
Hemolysis/Turbidity detection-quantitation no/yes
How labs get LOINC codes for reagent kits upon request
Interface supplied by instrument vendor no
Interfaces in active user sites for: all major LIS vendors
Lag time during which hypercoagulable sample not detected yes (PT: <7 seconds; PTT: <15 seconds)
List price CA-620: $42,000; CA-660: $55,000
LOINC codes transmitted with all results no
Maximum time same lot number of reagents can be used 12 months
Measures No. of tests remaining/Short sample detection yes/yes
Methodologies supported clot detection, optical; turbidimetric; chromogenic, immunologic
Minimum sample volume aspirated precisely at one time 5 µL
Model type benchtop
Modem servicing no
Multiple reagent configurations supported yes
Number of different assays programmed and calib. at one time 7
Number of different measured assays onboard simultaneously 5
Number of reagent containers onboard at once/Tests per container/Reagents refrigerated onboard 11/up to 200/yes (15°C)
Number of units installed in U.S./Outside U.S.
Number of user-definable (open) channels 7
Of those defined, number active simultaneously 5
Onboard maintenance records no
Onboard test automatic inventory yes
Operates on whole blood or spun plasma spun plasma
Operational type continuous random access
Operator must load sep. reagent pack per specimen/Test run no/no
Primary tube sampling supported/Pierces caps on primary tubes yes (2.7–5.0 mL)/no
Producer Siemens Healthcare Diagnostics
Read time extended for prolonged clotting times yes, selectable on operator menus
Reagent barcode reading capability no
Reagent type open reagent system (reconstituted manually)
Reagents, consumables loaded without interrupting testing yes
Results transferred to LIS as soon as test time complete yes
Same capabilities when third-party reagent used yes
Sample barcode reading capability yes
Sample handling system 10-tube position sample rack
Standard specimen volume required to run PT or PTT/ Factor VIII activity 50 µL/5 µL
Stat time to complete all analytes/Throughput per hour for Factor VIII activity assay
Stat time to complete all analytes/Throughput per hour for Fibrinogen 7 minutes/60 results
Stat time to complete all analytes/Throughput per hour for PT alone 7 minutes/60 results
Stat time to complete all analytes/Throughput per hour for PT, PTT 8 minutes/48 results
Supports direct-from-track sampling no
Tests in development but not yet submitted
Time delay from ordering stat to aspiration of sample 2 minutes
Time required for maintenance by lab personnel daily: <10 minutes; quarterly: <5 minutes
User can set different-than-standard Incubation times/Reading times yes/yes
User can set different-than-standard No. and sources of reagent yes
User can set different-than-standard Reagent volumes/Sample volumes yes/yes
Walkaway capacity: Number of specimens/Number of tests 18537

;